ANGIODYNAMICS ANGIODYNAMICS; DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION Back to Search Results
Model Number 45754
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/06/2023
Type of Reportable Event Malfunction
Event or Problem Description
An angiodynamics territory manager reported an issue with a mini stick max 5f x 10cm std.018 ss/ss echo 2.75" pg.During a procedure, part of the micropuncture seared off, inside of the patient, during withdrawal.The fragment did not require the use of a snare to remove.The procedure was completed with another same device and there was no report of patient harm.Multiple attempts have been made to obtain further details; however, this has been unsuccessful.
 
Additional Manufacturer Narrative
The device is not available to be returned to the manufacturer for evaluation.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Additional Manufacturer Narrative
The customer's reported complaint description of guidewire tip detached was not confirmed since no complaint sample was returned.Without receiving the complaint sample for evaluation, we are unable to definitively determine a root cause for this incident.The guidewire device is supplied to angiodynamics by the supplier/manufacturer heraeus medical.(b)(4) was sent to heraeus for dhr review of the reported lot number.(b)(4)response stated that no deviations or nonconformances related to the [tip detached] event description were identified in the record review.There is no evidence to confirm the issue is due to the manufacturing process.This event is considered a non-manufacturing-related issue.A potential root cause for this type of guidewire tip detached within the vasculature failure mode is end user technique in pulling guidewire back against the needle tip bevel.This is cautioned against in device directions for use.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: direction for use (16600601-01) is provided with this mini-stick device.The failure mode of device fracture and embolization is cautioned as potential adverse event.No sample was returned for evaluation so it cannot be determined if device was used in a manner inconsistent with labeling.Adverse events guidewire shearing, fracture or embolization.Operational instructions 1.Prior to insertion, flush all components with saline or heparinized/saline.2.Gain percutaneous access with the 21 g (0.9 mm) vascular introducer needle.3.Advance the 0.018 inch (0.46 mm) guidewire through the needle.Caution: the guidewire should not be withdrawn through the introducer needle.If the guidewire must be withdrawn while still inside the needle, simultaneously remove both the needle and the wire as a unit to prevent the needle from damaging the guidewire.4.Withdraw the introducer needle, leaving the 0.018 inch (0.46 mm) guidewire in place.5.Advance the coaxial assembly over the 0.018 inch (0.46 mm) guidewire.6.Simultaneously remove the dilator and 0.018 inch (0.46 mm) guidewire, while holding the sheath portion of the assembly in place.7.Advance a 0.035 inch (0.89 mm) or 0.038 inch (0.97 mm) guidewire into the sheath.8.Remove the sheath, leaving the 0.035 inch (0.89 mm) or 0.038 inch (0.97 mm) wire.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Additional Manufacturer Narrative
**udi related data quality updates only** the information in this report has been corrected to match the data submitted in global unique device identification database (gudid).Reference (b)(4).Corrections: d4: model #, udi.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIODYNAMICS
Common Device Name
DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
Manufacturer (Section G)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key18255296
Report Number1317056-2023-00169
Device Sequence Number7108170
Product Code DRE
UDI-Device IdentifierH965457540
UDI-PublicH965457540
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K132713
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/19/2023
Device Model Number45754
Device Catalogue Number45754
Device Lot Number5787808
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/06/2023
Supplement Date Received by Manufacturer02/20/2024
07/10/2024
Initial Report FDA Received Date12/04/2023
Supplement Report FDA Received Date02/27/2024
07/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-