| Catalog Number |
BSTW1GG625015 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
In this event it is reported that waveone gold glider 015-02/25mm blister 6 file broke during use.The broken part could not be retrieved.No injury from alleged event.
|
| |
|
Additional Manufacturer Narrative
|
|
Summary: returned protaper ultimate finishing file f1 21mm is broken at the tip of the active part (fatigue).No material defect was found during analysis of the rupture pattern.No unused file is available for evaluation.The batch number is unknown, dhr cannot be reviewed.Root causes are not identified.We will track this kind of event and monitor the trend.
|
| |
|