| Catalog Number |
EVLTPVAK |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported device has been returned to the manufacturer and an investigation into the root cause for event is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
|
| |
|
Event or Problem Description
|
|
An end user reported an issue with a pvak -- 400 micron perforator and accessory vein ablation kit.While unpacking the device, the fiber was found to be "bent, looking almost broken, near the connector." there was no patient involvement and there was no evident damage to the packaging of the device.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer's reported complaint description of the fiber was fractured was confirmed based on visual inspection of the returned fiber complaint sample.The od of the fiber shaft near the fracture location was confirmed to meet device specification.The likely root cause of the fiber fracture is handling damage but when and how this occurred cannot be definitively determined.The fiber tip is packaged such that the tip of the fiber is located under the coil wrap.The coil wrap was still in place for this complaint sample.A potential contributing factor for the fiber fracture is handling damage during the unpacking process.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: the directions for use (14601411-01) which is supplied to the end user with the reported catalog number contains the following statement: warning contents supplied sterile using an ethylene oxide (eo) process.Do not use if sterile barrier is damaged.If damage is found, call your sales representative.Inspect prior to use to verify that no damage has occurred during shipping.Intended use the venacure evlt 400 m perforator and accessory vein ablation kit is intended for use in the treatment of superficial vein reflux of the greater saphenous vein associated with varicosities.The venacure evlt 400 m perforator and accessory vein ablation kit is indicated for treatment of incompetence and reflux of superficial veins in the lower extremity, and for treatment of incompetent (i.E.Refluxing) perforator veins (ipvs).Equipment handling requirements venacure evlt 400 m perforator and accessory vein ablation kit: using sterile technique open the pack, place all contents into sterile field and inspect for damage.Do not use if any component is damaged.If damage is present, contact customer service or your local representative.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference pr(b)(4).
|
| |
|
Search Alerts/Recalls
|