• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS 400 MICRON FIBER PROCEDURE KIT; VENACURE ENDOVENOUS LASER TREATMENT FIBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS 400 MICRON FIBER PROCEDURE KIT; VENACURE ENDOVENOUS LASER TREATMENT FIBER Back to Search Results
Catalog Number EVLTPVAK
Medical Device Problem Code Component Missing (2306)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/27/2023
Type of Reportable Event Malfunction
Event or Problem Description
A nurse practitioner reported the following issue with a pvak -- 400 micron fiber procedure kit: when opening a procedure kit, it was noted the clear end of the lasser was missing.The device was set aside and a new of the same device was opened in order to continue with the procedure.
 
Additional Manufacturer Narrative
The reported device has been returned to the manufacturer and an investigation into the root cause for event is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Additional Manufacturer Narrative
The customer's reported complaint description of the clear fiber tip was fractured and detached was confirmed based on visual inspection of the returned fiber complaint sample.A likely root cause of the fiber fracture is handling damage but when and how this occurred cannot be definitively determined.The fiber tip is packaged such that the tip of the fiber is located under the coil wrap.A potential contributing factor for the fiber fracture is the coil wrap and/or the process of end user removing the coil wrap from the fiber during the unpacking process.Manufacturing personnel 100% visually inspect devices during the packaging process.This type of fiber kink/detached damage would be noticed prior to shipment.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: the directions for use (14601411-01) which is supplied to the end user with the reported catalog number contains the following statement: warning: contents supplied sterile using an ethylene oxide (eo) process.Do not use if sterile barrier is damaged.If damage is found, call your sales representative.Inspect prior to use to verify that no damage has occurred during shipping.Intended use: the venacure evlt 400 m perforator and accessory vein ablation kit is intended for use in the treatment of superficial vein reflux of the greater saphenous vein associated with varicosities.The venacure evlt 400 m perforator and accessory vein ablation kit is indicated for treatment of incompetence and reflux of superficial veins in the lower extremity, and for treatment of incompetent (i.E.Refluxing) perforator veins (ipvs).Equipment handling requirements: venacure evlt 400 m perforator and accessory vein ablation kit: using sterile technique open the pack, place all contents into sterile field and inspect for damage.Do not use if any component is damaged.If damage is present, contact customer service or your local representative.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
400 MICRON FIBER PROCEDURE KIT
Common Device Name
VENACURE ENDOVENOUS LASER TREATMENT FIBER
Manufacturer (Section D)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer (Section G)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key18256641
Report Number1319211-2023-00079
Device Sequence Number20612108
Product Code GEX
UDI-Device IdentifierH787EVLTPVAK5
UDI-PublicH787EVLTPVAK5
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K041957
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Catalogue NumberEVLTPVAK
Device Lot Number5791327
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2023
Supplement Date Received by Manufacturer03/08/2024
Initial Report FDA Received Date12/04/2023
Supplement Report FDA Received Date03/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/26/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-