| Model Number |
11007641 |
| Medical Device Problem Codes |
No Display/Image (1183); No Device Output (1435)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Siemens is conducting a thorough investigation of the reported event.A supplement report will be filed if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
Siemens became aware of a malfunction that occurred while operating the sensis vibe hemo system.Prior to an emergency patient procedure, there was no display output.The failure caused a long-term delay of the emergency procedure.We have no indications of any adverse effects on the health status of the involved patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was completed by our experts.The on-site service intervention identified a loose video adapter cable.Neither a defect in the cable nor why or how the cable became loose could be determined.As a preventive measure, the cable was replaced by siemens local service.The occurrence rate of the aforementioned error pattern was checked.A possible error accumulation or even a systematic error, which leads to a corrective action of the installed base, could not be determined by the investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
Purpose of this supplemental report is to correct section d4 ¿primary udi number¿ per request by the mdr data systems team (cdrh) dated june 11, 2024.The primary udi number now contains the di and pi numbers as requested.
|
| |
|