| Catalog Number |
EVLTPVAK |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported device has yet to be returned to the manufacturer and an investigation into the root cause for event is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
|
| |
|
Event or Problem Description
|
|
A medical assistant reported an issue with a pvak -- 400 micron perforator and accessory vein ablation kit.It was reported that the fiber was found to be bent, upon opening the device.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer's reported complaint description of the fiber fractured upon opening the device was confirmed based on visual inspection of the returned fiber complaint sample.The coil wrap was almost completely removed and fiber was fractured near the coil wrap location.The fiber od was confirmed to meet product specification and there was no report that fiber was fractured when removed from the packaging.The likely root cause of the fiber fracture is handling damage by end user during the coil wrap removal process.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: the directions for use (14601411-01) which is supplied to the end user with the reported catalog number contains the following statement: warning contents supplied sterile using an ethylene oxide (eo) process.Do not use if sterile barrier is damaged.If damage is found, call your sales representative.Inspect prior to use to verify that no damage has occurred during shipping.Intended use the venacure evlt 400]m perforator and accessory vein ablation kit is intended for use in the treatment of superficial vein reflux of the greater saphenous vein associated with varicosities.The venacure evlt 400 m perforator and accessory vein ablation kit is indicated for treatment of incompetence and reflux of superficial veins in the lower extremity, and for treatment of incompetent (i.E.Refluxing) perforator veins (ipvs).Equipment handling requirements venacure evlt 400 m perforator and accessory vein ablation kit: using sterile technique open the pack, place all contents into sterile field and inspect for damage.Do not use if any component is damaged.If damage is present, contact customer service or your local representative.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference pr(b)(4).
|
| |
|
Search Alerts/Recalls
|