|
The report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 8 months since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of foreign material on the device could not be identified.However, it¿s likely the cause is related to appropriate reprocessing not being performed according to the instruction manual.The event (foreign material on the device) can be prevented by following the instructions for use which state: ¿ before use, confirm that the distal tip of the forceps is not corroded, dented, or discolored, and that the cups can be opened and closed smoothly.Using forceps that are damaged or not working properly may result in one or more of its components falling off inside the patient.If a component falls off the distal end, or if operation of the cups suddenly becomes more difficult, immediately stop the procedure.Carefully withdraw the forceps together with the endoscope to avoid causing injury within the body cavity.Retrieve any parts that have fallen off inside the patient using another grasping forceps.¿ reprocess the instrument immediately after use by immersing it in a neutral, low-foaming, medical-grade detergent solution, then following the cleaning and sterilization procedures in this chapter.Failure to reprocess the instrument immediately after use, or using another type of detergent may cause corrosion at the instrument¿s distal end.This could cause components to fall off during use and may interfere with operation of the cups.Olympus will continue to monitor field performance for this device.
|