ABBOTT VASCULAR RX ACCULINK CAROTID STENT SYSTEM Back to Search Results
Catalog Number 1011343-40
Medical Device Problem Codes Break (1069); Mechanical Jam (2983); Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/18/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned for analysis.The reported mechanical jam and the reported activation failure were unable to be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no damage noted to the device during the inspection prior to use, the investigation determined the reported/noted difficulties appear to be related to circumstances of the procedure as it is likely that during advancement interaction with the heavily calcified, heavily tortuous and 95% stenosed anatomy resulted in preventing the shaft lumens from moving freely; thus resulting in resistance with the slider and the difficulty deploying the stent.Manipulation of the device ultimately resulted in the noted hypotube separation contributing to the reported difficulties.There is no indication of a product quality issue with respect to design, manufacture or labeling.
 
Event or Problem Description
It was reported that the procedure was to a 95% stenosed lesion in the internal carotid artery (ica) with heavy calcification and heavy tortuosity.The 6-8x40mm acculink self expanding stent system (sess) was advanced; however, the stent failed to deploy because the thumbslide was stuck.Another acculink sess was used in the procedure.There was no adverse patient effect and no clinically significant delay reported in the procedure.Return device analysis found that the hypotube was separated inside the handle.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX ACCULINK CAROTID STENT SYSTEM
Common Device Name
CAROTID STENT
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18259135
Report Number2024168-2023-13413
Device Sequence Number7859432
Product Code NIM
UDI-Device Identifier08717648076374
UDI-Public08717648076374
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P040012
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Catalogue Number1011343-40
Device Lot Number2081961
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/09/2023
Initial Report FDA Received Date12/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-