|
The device was returned to olympus for evaluation.The device evaluation found the following: due to a pinhole on the bending section cover, water tightness was lost; due to a dent on the distal end, water tightness was lost; due to damage to the channel tube, the channel cleaning brush could not be inserted smoothly; the bending tube had a dent; scratches were found on multiple locations on the device; and the connecting tube has a dent.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device was returned, and an evaluation was completed.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue were identified.Based on the results of the investigation, it is likely that stress of repeated use, external factors, or handling of the device caused the peeling to occur.However, a definitive root cause could not be determined.The following information is stated in the instructions for use (ifu): 3.2 ¿inspection of the endoscope¿.5.¿inspect the external surface of the entire insertion tube including the bending section and the distal end for dents, bulges, swelling, peeling, scratching, holes, sagging, transformation, bends, adhesion of foreign body, dropout of parts, any protruding objects or other irregularities.¿.Olympus will continue to monitor field performance for this device.
|