| Catalog Number |
09P31-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Codes |
Cellulitis (1768); Fever (1858)
|
| Date of Event |
11/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2023, abbott point of care (apoc) was contacted by a customer regarding i-stat chem8+ cartridge that yielded discrepant creatinine results on a 73 year old female patient with fever, cellulitis post mastectomy.There was no additional patient information at the time of this report.Return product is not available for investigation.Method date collected tested results sample i-stat on (b)(6) 2023 08:40 08:48 576 umol/l b i-stat on (b)(6) 2023 09:00 09:06 605 umol/l c lab (b)(6) 2023 ni 14:55 55 umol/l ni at this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggested that product was not performing within the variability.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: # (b)(4).The investigation was completed on 15-jan-2024.A review of the device history record (dhr) confirmed the cartridge lot met finished goods (fg) release criteria.Retained cartridge testing met the acceptance criteria outlined in appendix 1 of q04.01.003 rev.Al (product complaint level 2 and level 3 investigation procedure).No deficiency has been identified for chem8+ cartridge lot h23191.
|
| |
|
Event or Problem Description
|
|
Na.
|
| |
|