| Catalog Number |
09P31-26 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
Heart Failure/Congestive Heart Failure (4446)
|
| Date of Event |
11/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 30-nov-2023, abbott point of care (apoc) was contacted by a customer regarding i-stat chem8+ cartridge that yielded discrepant potassium results on a 63 year old male patient on a routine visit.There was no additional patient information at the time of this report.Return product is available for investigation.Method date collected tested result sample i-stat (b)(6) -2023 16:27 16:31 >9 mmol/l a lab (b)(6) -2023 16:39 17:22 4.1 mmol/l b at this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggested that product was not performing within the variability.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: (b)(4).The investigation was completed on 17-jan-2024.A review of the device history record (dhr) confirmed the cartridge lot met finished goods release criteria.Retained and returned cartridge testing met the acceptance criteria found in q04.01.003 rev.Al, appendix 1 - product complaint level 2 and level 3 investigation procedure.No deficiency has been determined for chem8+ cartridge lot h23233.
|
| |
|
Event or Problem Description
|
|
Na.
|
| |
|