• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT CHEM8+ CARTRIDGE; CHEM8+ CARTRIDGE,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT CHEM8+ CARTRIDGE; CHEM8+ CARTRIDGE, Back to Search Results
Catalog Number 09P31-26
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code Heart Failure/Congestive Heart Failure (4446)
Date of Event 11/29/2023
Type of Reportable Event Malfunction
Event or Problem Description
On 30-nov-2023, abbott point of care (apoc) was contacted by a customer regarding i-stat chem8+ cartridge that yielded discrepant potassium results on a 63 year old male patient on a routine visit.There was no additional patient information at the time of this report.Return product is available for investigation.Method date collected tested result sample i-stat (b)(6) -2023 16:27 16:31 >9 mmol/l a lab (b)(6) -2023 16:39 17:22 4.1 mmol/l b at this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggested that product was not performing within the variability.The investigation is underway.
 
Additional Manufacturer Narrative
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Additional Manufacturer Narrative
Apoc incident: (b)(4).The investigation was completed on 17-jan-2024.A review of the device history record (dhr) confirmed the cartridge lot met finished goods release criteria.Retained and returned cartridge testing met the acceptance criteria found in q04.01.003 rev.Al, appendix 1 - product complaint level 2 and level 3 investigation procedure.No deficiency has been determined for chem8+ cartridge lot h23233.
 
Event or Problem Description
Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT CHEM8+ CARTRIDGE
Common Device Name
CHEM8+ CARTRIDGE,
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
ABBOTT POINT OF CARE CANADA LTD.
185 corkstown road
ottawa, ontario K2H 8 V4
CA   K2H 8V4
Manufacturer Contact
linda maczuszenko
185 corkstown road
ottawa, ontario, NJ K2H 8-V4
CA   K2H 8V4
6136123525
MDR Report Key18266110
Report Number2245578-2023-00172
Device Sequence Number7691176
Product Code JGS
UDI-Device Identifier10054749005519
UDI-Public10054749005519
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K183688
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/17/2024
Device Catalogue Number09P31-26
Device Lot NumberH23233
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/30/2023
Supplement Date Received by Manufacturer01/17/2024
Initial Report FDA Received Date12/05/2023
Supplement Report FDA Received Date01/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/21/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age63 YR
Patient SexMale
-
-