ETHICON INC. SFX SPI PDS+ UNI VIO 6IN 2-0 S/A SH; SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE
|
Back to Search Results |
|
| Catalog Number |
SXPP1B415 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 12/5/2023 h6 component code: g07002 - device not returned this report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: *device follow up to date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The following information was received: was surgery delayed due to the reported event? --> no, was procedure successfully completed? --> yes, were fragments generated? --> no, if yes, were they removed easily without additional intervention? --> unknown, patient status/ outcome / consequences --> no, patient consequence description/was there a clinical outcome experienced by the patient (infection, inflammation, etc.)?--> nil consequence, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: --> unknown, is the patient part of a clinical study --> unknown related reports: 2210968-2023-09455.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a lap hysterectomy procedure on (b)(6) 2023 and suture was used.During the surgery, the suture passed through 5mm trocar and needle snapped off from suture.Suture held by needle holder at about mid point of suture when passing through and normally passes through easily.The suture snapped on 2 x sutures from the same box of suture however only one suture available for return.There were no patient consequences reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 investigation summary: the product was returned for evaluation.Visual inspection was conducted on the returned device.The returned sample revealed that it was returned, a detached needle and suture outside the packaging that pertain to the product code sxpp1b415.Upon visual inspection of the returned sample, the swage and attachment area were noted to be as expected.The barrel hole of the needle was examined under magnification, and no suture remnant was noted.The suture was examined, and a mark of correct insertion was noted in one of the extreme of the suture; however, the force used to detach the needle from the suture could not be determined.After further evaluation crimping marks near the swage area were observed on the suture possibly caused by a surgical instrument.The condition of the sample received indicates improper handling of the device.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of surgical instruments, such as needle holders and forceps.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.No conclusion could be reached as to what caused the reported complaint.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.
|
| |
|