• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA LINER: VERSAFITCUP CC TRIO FLAT PE HC LINER Ø 32 / C; HIP SHELL LINER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA LINER: VERSAFITCUP CC TRIO FLAT PE HC LINER Ø 32 / C; HIP SHELL LINER Back to Search Results
Catalog Number 01.26.3239HCT
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Unequal Limb Length (4534)
Date of Event 11/09/2023
Type of Reportable Event Serious Injury
Event or Problem Description
The patient came in reporting pain and instability due to a leg length discrepancy.About 1 year and 11 months after the primary surgery, the surgeon revised the head, cup and liner.The surgery was completed successfully.
 
Additional Manufacturer Narrative
Batch review performed on 16 november 2023: lot 2111759: (b)(4) items manufactured and released on 14-sep-2021.Expiration date: 2026-aug-31.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Additional devices involved batch review performed on 16 november 2023.Ball heads: mectacer 01.29.206 biolox delta ceramic ball head 12/14 ø 32 size l +4 (k112115) lot 2009397: (b)(4) items manufactured and released on 19-jan-2021.Expiration date: 2026-jan-07.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Cup: versafitcup cc trio 01.26.45.0046 acetabular shell cc trio ø 46 (k103352) lot 2104085: (b)(4) items manufactured and released on 15-jul-2021.Expiration date: 2026-jul-04.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINER: VERSAFITCUP CC TRIO FLAT PE HC LINER Ø 32 / C
Common Device Name
HIP SHELL LINER
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key18266626
Report Number3005180920-2023-00962
Device Sequence Number15426544
Product Code LZO
UDI-Device Identifier07630030807541
UDI-Public07630030807541
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K120531
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2026
Device Catalogue Number01.26.3239HCT
Device Lot Number2111759
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2023
Initial Report FDA Received Date12/05/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient SexFemale
-
-