Catalog Number 55625800
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/12/2023
Type of Reportable Event Malfunction
Event or Problem Description
As reported by the health authority, the seal of an envoy 5f, mpd 100 cm (55625800, 30885295) was found untight when preparing for an unknown operation.Th physician changed to another device to complete the surgery.No additional information is available.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Section e1.Initial reporter phone: (b)(6).The device was discarded; therefore, no further investigation can be performed.A manufacturing record evaluation was performed for the finished device 30885295 number, and no non-conformances related to the malfunction were identified.With the information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the manufacturing record evaluation, there is no indication that the event is related to the device manufacturing process.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
**udi related data quality updates only**.Correction to the initial mdr, incomplete udi was provided.Section d4.Primary udi number has been updated with full udi (b)(4).If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.No: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENVOY 5F, MPD 100 CM
Common Device Name
PERCUTANEOUS CATHETER
Manufacturer (Section D)
CERENOVUS, INC.
6303 waterford district drive
suites 215 & 315
miami FL 33126
Manufacturer (Section G)
CERENOVUS INC. (JUAREZ)
circuito interior norte #1820
parque industrial salvarcar
juarez chihuahua 32574
MX   32574
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18267011
Report Number3007628272-2023-00002
Device Sequence Number19854516
Product Code DQY
UDI-Device Identifier10886704027959
UDI-Public(01)10886704027959(11)220811(17)250731(10)30885295
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K000715
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Catalogue Number55625800
Device Lot Number30885295
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/22/2023
Supplement Date Received by Manufacturer07/10/2024
Initial Report FDA Received Date12/05/2023
Supplement Report FDA Received Date07/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-