BOSTON SCIENTIFIC COSTA RICA TRANSEND EX 14 FLOPPY 205CM; WIRE, GUIDE, CATHETER Back to Search Results
Catalog Number M001468070
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3 other text : the device is not availabel to the manufacturer.
 
Event or Problem Description
It was reported that during one middle cerebral artery mca stenosis case, when using subject guidewire to build access, friction was encountered when advancing the subject guidewire into microcatheter.Withdrew the guidewire and found coating of the tip was peeled off.The procedure was completed successfully by replacing the subject device.No clinical consequences were reported to the patient due to this event.
 
Event or Problem Description
It was reported that during one middle cerebral artery mca stenosis case, when using subject guidewire to build access, friction was encountered when advancing the subject guidewire into microcatheter.Withdrew the guidewire and found coating of the tip was peeled off.The procedure was completed successfully by replacing the subject device.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
B1 adverse event/product problem - corrected - no product problem.H1 type of reportable event - corrected - no malfunction.Based on the results of the device history record (dhr) review, there is no indication that the device, labeling or packaging failed to meet its specifications when released.During visual/microscopic inspection, the guidewire tip was seen to be kinked/bent to the distal end.The hydrophilic coating to the guidewire tip was seen to have been used/hydrated, however no damage noted.No other anomalies were noted.There were no anomalies/peeling noted to either the distal hydrophilic coating or the proximal ptfe coating.During functional inspection, the guidewire coating was hydrated; no anomalies were noted.The guidewire was advanced through a demonstration microcatheter without difficulties.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed based on analysis.The returned device failed to meet specifications when received for complaint investigation, based on the damage noted.It was reported that when using subject guidewire to build access friction was encountered when advancing the guidewire into microcatheter.Withdrew the guidewire and found tip of coating was peeled off.The device was received and the there was no visual damage noted to the coatings, the wire was able to be advanced through a demonstration microcatheter without difficulty.There was a slight kink to the distal end which likely occurred during advancing the guidewire into the microcatheter.An assignable cause of not conformed will be assigned to the reported 'guidewire difficult to advance' and 'hydrophilic coating peeling', as the issue was not confirmed during analysis.It is probable that the guidewire distal tip was kinked during use, therefore an assignable cause of procedural factors will be assigned to the analyzed 'guidewire kinked/bent'.The manufacturer has reviewed all information and determined this event no longer meets the requirement of the reportable event for the device in question.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSEND EX 14 FLOPPY 205CM
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC COSTA RICA
coyol facility, 2546 first st.
propark free zone
alajuela 20101
CS  20101
Manufacturer (Section G)
BOSTON SCIENTIFIC COSTA RICA
coyol facility, 2546 first st.
propark free zone
alajuela 20101
CS   20101
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key18267399
Report Number3008853977-2023-00051
Device Sequence Number7900190
Product Code DQX
UDI-Device Identifier08714729326250
UDI-Public08714729326250
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K944677
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/21/2024
Device Catalogue NumberM001468070
Device Lot Number30583514
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/16/2023
Supplement Date Received by Manufacturer01/31/2024
Initial Report FDA Received Date12/05/2023
Supplement Report FDA Received Date02/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
XT27 MICROCATHETER (STRYKER).
Patient Age80 YR
Patient SexFemale
Patient RaceAsian
-
-