| Model Number |
HMT2100 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/26/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No additional information has been provided.A supplemental report will be submitted once the manufacturer's investigation is complete.
|
| |
|
Event or Problem Description
|
|
Related manufacturer report number: 2916596-2023-08312.It was reported that on 26nov2023, the system controller was connected to the power module via the black and white power cables.The heartmate touch was also connected to the power module and displayed the device parameters.At around 18:30, the data stopped displaying on the heartmate touch.The connection between the heartmate touch and system controller was retried but did not work.The heartmate touch was unable to recognize that the system controller was connected.A new heartmate touch was tried with the same controller and patient cable but nothing changed.The patient cable was then exchanged but the issue persisted.The system monitor was then connected, and the parameters were able to be displayed.The next day, the heartmate touch wireless adapter was exchanged, and the issue resolved.The original adapter was tried again, to see if it would still work, and it functioned as intended.Both adaptors appeared to work properly.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation conclusion: the reported event of a loss of communication between the system controller and heartmate touch was not confirmed.It was reported that changing the wireless adapter resolved the issue; however, the exchanged wireless adapter was then tested again and was found to be working.Additional information provided stated that there have been no further connection issues and no damage to the system controller or wireless adapter was observed.It was also reported that no product will be returned as it is still in use.The root cause of the reported event could not be conclusively determined through this analysis.The wireless adapter is manufactured by a third party who maintains the device history records.Heartmate 3 instructions for use (ifu) chapter 4 ¿heartmate touch communication system¿ and the heartmate touch communication system quick start guide under ¿how to use the heartmate touch communication system¿ explain the operation of the heartmate touch system, including how to set up the system and connect the tablet to the wireless adapter and system controller and how to use the heartmate touch app.These documents also warn that the heartmate touch system may be interfered with by other equipment, even if that other equipment complies with cispr emission requirements.Heartmate 3 ifu chapter 7 ¿alarms and troubleshooting¿ and the heartmate touch communication system quick start guide under ¿troubleshooting¿ cover all alarms (visual and audible), including the disconnected system controller - heartmate touch app error message.Heartmate 3 ifu section 8 "equipment storage and care" describes how to care for and clean all equipment.Heartmate 3 ifu section f, both entitled ¿safety checklists¿, provide checklists to assist the patient in performing routine maintenance of the heartmate 3 lvad.This section also informs the user to replace any equipment or system component that appears damaged or worn.Heartmate 3 patient handbook cautions users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
|
| |
|
Search Alerts/Recalls
|