| Catalog Number |
22-4036 |
| Medical Device Problem Codes |
Break (1069); Component Missing (2306); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3: the sample is under evaluation by the manufacturing site.
|
| |
|
Event or Problem Description
|
|
It was reported that during a shoulder arthroscopy, during the assembly of the firstpass needle to the clamp, it was observed that the plate (step 1) was disassembled from the attachment and was missing the assembly hook at the tip.The procedure was completed with a non-significant delay using a s+n back up device.No further complications were reported.Preliminary results of investigation have concluded that was found that the hook on the tip has been broken off.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: the reported device was received for evaluation.A visual inspection revealed the device was returned in original packaging with the batch number of the complaint on the label.The suture capture was returned off of the insertion tool.There is bio debris on the suture capture.The hook on the tip has been broken off.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with unintended use of the device.Factors that could have contributed to the failure include an application of unintended inappropriate or excessive force to the device.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|