Catalog Number 71655136
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Failure of Implant (1924); Pain (1994)
Date of Event 10/25/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, after an internal fixation surgery performed on (b)(6) 2023, the patient indicated that on (b)(6) 2023 the meta-nail tibial 10mm x 36cm that was implanted broke in half.The surgeon who implanted the device drilled holes in it to secure it with more hardware, but it still broke.On (b)(6) 2023 a revision surgery was performed in order to replace the broken nail.The current health status of the patient is good.
 
Additional Manufacturer Narrative
H10; additional information: updated section b5 and h6 (clinical code).
 
Event or Problem Description
It was reported that, after an internal fixation surgery performed on (b)(6) 2023, the patient indicated that on (b)(6) 2023 the meta-nail tibial 10mm x 36cm that was implanted broke in half casing pain and angulation of the site.The surgeon who implanted the device drilled holes in it to secure it with more hardware, but it still broke.On (b)(6) 2023 a revision surgery was performed in order to replace the broken nail.The current health status of the patient is good.
 
Additional Manufacturer Narrative
Section h3, h6: the device was not returned for evaluation.However, the photographs were reviewed, and revealed that the nail fractured into two pieces.The clinical/medical investigation concluded that, based on the information provided, it can be concluded that the reported meta-tan nail breakage is not indicative of a malperformance of the smith and nephew device.The customization of the implant (drilling additional holes for screw mid shaft) coupled with the patient¿s persistent fractures and reported activity are likely contributing factors to the reported event.The surgeon confirmed by reporting ¿additional risks for him are that he was putting a huge number of cycles walking on the nail and there was a segment that was slower than normal in consolidating.¿ instructions for use for fracture fixation devices, notes "use extreme care in handling and storing implant components.Cutting, bending, or scratching the surface of metal components can cause internal stresses which significantly reduce the strength and fatigue resistance.It was further reported per the surgeon " the nail broke before we would have expected complete healing either way.¿ the patient¿s current health status is reportedly ¿good.¿ the impact is the revision of this nail and extended rehabilitation.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for intramedullary nail system revealed that cracking or fracture of the implant components have been identified in possible adverse effects section.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.According with material specification, the quality and manufacture of standard grade titanium-6 aluminum-4 vanadium alloy shall be controlled.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include size selected, surgical technique, postoperative care, patient anatomy, abnormal loading of limb and/or excessive forces.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.H6: health effect - clinical code.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
META-NAIL TIBIAL 10MM X 36CM
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18267466
Report Number1020279-2023-02378
Device Sequence Number14473907
Product Code JDS
UDI-Device Identifier03596010585585
UDI-Public03596010585585
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K210980
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71655136
Device Lot Number21KT55678
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/13/2023
Supplement Date Received by Manufacturer11/27/2023
12/26/2023
Initial Report FDA Received Date12/05/2023
Supplement Report FDA Received Date12/08/2023
12/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Other;
Patient Age46 YR
Patient SexMale
Patient Weight86 KG
-
-