| Catalog Number |
1013319J |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Separation (1562); Stretched (1601); Device Damaged by Another Device (2915)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
11/13/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents/complaints from this lot.The investigation determined the reported difficulties appear to be related to the operational context of the procedure.Additionally, the separated portion of the distal tip remains free floating in the artery.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat the left anterior descending (lad) and diagonal coronary artery bifurcation with high calcification.The versaturn guide wire was placed in the diagonal artery and a non-abbott guide wire was placed in the lad.A non-compliant balloon was used due to the high calcification and then a stent was then implanted in the lad.This trapped the versaturn guide wire and the distal tip separated and the wire unraveled.The distal tip remains free floating in the artery and no dilatation of the diagonal artery was able to be performed.The patient was doing well at the end of the procedure.No additional information was provided.
|
| |
|