• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR HT VERSATURN GUIDE WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR HT VERSATURN GUIDE WIRE Back to Search Results
Catalog Number 1013319J
Medical Device Problem Codes Difficult to Remove (1528); Material Separation (1562); Stretched (1601); Device Damaged by Another Device (2915)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 11/13/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents/complaints from this lot.The investigation determined the reported difficulties appear to be related to the operational context of the procedure.Additionally, the separated portion of the distal tip remains free floating in the artery.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Event or Problem Description
It was reported that the procedure was to treat the left anterior descending (lad) and diagonal coronary artery bifurcation with high calcification.The versaturn guide wire was placed in the diagonal artery and a non-abbott guide wire was placed in the lad.A non-compliant balloon was used due to the high calcification and then a stent was then implanted in the lad.This trapped the versaturn guide wire and the distal tip separated and the wire unraveled.The distal tip remains free floating in the artery and no dilatation of the diagonal artery was able to be performed.The patient was doing well at the end of the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HT VERSATURN GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005737652
road no.2 km 58.0 cruce dávila
barceloneta PR 00617
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18267910
Report Number2024168-2023-13481
Device Sequence Number11751197
Product Code DQX
UDI-Device Identifier08717648198496
UDI-Public08717648198496
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K141782
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Catalogue Number1013319J
Device Lot Number3091471
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/13/2023
Initial Report FDA Received Date12/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/14/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDE 7F MACH CLS 3.5; SION BLUE GUIDE WIRE
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-