| Model Number |
CSC100 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: 81 other - the customer mentioned thatt the unit/suspect device is not available for evaluation.Therefore, rootcause could not be determined.This report is for airlife¿ verso¿ airway access adapter.Please see (b)(4) for airlife closed suction catheter 14fr.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.H3 other text : device not available for evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported to vyaire medical that there was an stagnation of purulent secretions at the connector of csc100 adapter verso airway adult/ped 20/bx and csc114 catheter 14fr closed 50/c.There is no information on patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: g3, g6, h2, h3, h6 and h10.The dhr records has been thoroughly reviewed, and no abnormalities were detected during the inspection.The retained sample from the same production underwent testing under 80-200mmhg pressure, and its function is deemed acceptable.Upon reviewing the photos, it is observed that the secretions at the connector appear to be too thick.This thickness may lead to sticking on the valve of csc100, impacting the catheter's suction and potentially causing retention at the connector.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: d4:corrected model and unique identifier(udi).D4 was updated to indicate correct model and udi #.
|
| |
|