• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. FIBERTAG TIGHTROPE II ABS; NON-DEGRADABLE FIXATN FASTENER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. FIBERTAG TIGHTROPE II ABS; NON-DEGRADABLE FIXATN FASTENER Back to Search Results
Model Number FIBERTAG TIGHTROPE II ABS
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/07/2023
Type of Reportable Event Malfunction
Event or Problem Description
On 11/8/2023, it was reported by a sales representative via phone that an ar-1588tnt2 fibertag tightrope ii abs implant had an issue.When the needle was inserted into the quad tendon during graft preparation, it pulled away from the suture.The case was completed using another ar-1588tnt2 fibertag tightrope ii abs implant with no delay in the procedure and no fragments inside the patient.This was discovered during an acl reconstruction procedure on (b)(6) 2023, with no reported patient harm.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.The complaint allegation cannot be confirmed as the device was not received for evaluation, and no pictures of the failure were received.Based on the information provided, which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, it was concluded that the most likely cause of the reported failure is user error.Specifically, the error involved applying excessive force, most likely by pulling the needle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBERTAG TIGHTROPE II ABS
Common Device Name
NON-DEGRADABLE FIXATN FASTENER
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key18268507
Report Number1220246-2023-09181
Device Sequence Number3890745
Product Code MBI
UDI-Device Identifier00888867426696
UDI-Public00888867426696
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K202581
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 01/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFIBERTAG TIGHTROPE II ABS
Device Catalogue NumberAR-1588TNT2
Device Lot Number15139427
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2023
Supplement Date Received by Manufacturer12/26/2024
Initial Report FDA Received Date12/05/2023
Supplement Report FDA Received Date01/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/08/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-