| Catalog Number |
1400200-12 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.The investigation determined the reported complaint appears to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a lesion in the 90% stenosed, mildly calcified and mildly tortuous distal right coronary artery (drca).The balloon of a 2.0x12mm nc trek neo balloon dilatation catheter (bdc) was inflated for 5 seconds.During the first inflation at 6 atmospheres, the balloon ruptured.A 2.00 x 12mm mini trek bdc was advanced and the balloon inflated for 4 seconds to 5 atmospheres; however, also ruptured.Lastly a 2.0x20mm mini trek bdc was advanced and the balloon inflated for 4 seconds, the balloon ruptured at 5 atmospheres on the first inflation.The procedure was completed with a sapphire bdc without reported issue.There were no adverse patient effects and no clinically significant delay reported.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|