ABBOTT MEDICAL AMPLATZER VASCULAR PLUG II; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
|
Back to Search Results |
|
| Catalog Number |
9-AVP2-008 |
| Medical Device Problem Codes |
Off-Label Use (1494); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Obstruction/Occlusion (2422)
|
| Date of Event |
12/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
As reported through a literature article, "risk factors associated with device embolisation or malposition during transcatheter closure of patent ductus arteriosus", a 1 month old patient was born pre-term.A 6mm amplatzer vascular plug iv was chosen for implant to treat patent ductus arteriosus (pda) that was sized 4 mm.Then, the device had embolized and a decision was made to retrieve the device via transcatheter snare and replaced with an 8mm amplatzer vascular plug ii.After intervention procedure, it was reported the patient developed left pulmonary stenosis.The article concluded device embolisation complicates 4% of transcatheter duct closures, with need for surgery in one-fifth of them.Larger ducts with high pulmonary artery pressures in younger and smaller infants are more often associated with device embolisation.Tubular ducts are more prone for embolisation compared to usual conical ducts.Softer vascular plugs are often associated with embolisations.Intentional device undersizing to avoid vascular obstruction in small patients is a frequent risk factor for embolisation.Precise echocardiographic measurements, correct occluder choice, proper technique and additional care in patients with high pulmonary artery pressures are mandatory to minimise embolisations.A more comprehensive assessment could not be performed as the event was non-contemporaneously reported through a literature review and no device or individual patient information was received for analysis.Please note, per amplatzer vascular plug and vascular plug ii, instructions for use, artmt600034447, revision c, "the amplatzer¿ vascular plug and amplatzer¿ vascular plug ii are indicated for arterial and venous embolizations in the peripheral vasculature." this is considered as an off-label use of the device.However, it was unable to determine if the off-label use contributed to the reported incident.Therefore, a letter will not be provided to address the deviation from the ifu.
|
| |
|
Event or Problem Description
|
|
The article, ¿risk factors associated with device embolisation or malposition during transcatheter closure of patent ductus arteriosus¿, was reviewed.The article presented a case study of a 1 month old patient weighing 2.1 kg that was born pre-term.It was reported that on an unknown date, a 6mm amplatzer vascular plug iv was chosen for implant to treat patent ductus arteriosus (pda) that was sized 4 mm.It was then reported on an unknown date, the device had embolized and a decision was made to retrieve the device via transcatheter snare and replace with an 8mm amplatzer vascular plug ii.After intervention procedure, it was reported the patient developed left pulmonary stenosis.The article concluded device embolisation complicates 4% of transcatheter duct closures, with need for surgery in one-fifth of them.Larger ducts with high pulmonary artery pressures in younger and smaller infants are more often associated with device embolisation.Tubular ducts are more prone for embolisation compared to usual conical ducts.Softer vascular plugs are often associated with embolisations.Intentional device undersizing to avoid vascular obstruction in small patients is a frequent risk factor for embolisation.Precise echocardiographic measurements, correct occluder choice, proper technique and additional care in patients with high pulmonary artery pressures are mandatory to minimise embolisations.[the primary and secondary author was kothandam sivakumar, madras medical mission, 4a dr jj nagar, mogappair, chennai 600089, india, with corresponding email drkumarsiva@hotmail.Com].
|
| |
|
Search Alerts/Recalls
|
|
|