| Model Number |
ACU0T0 |
| Medical Device Problem Codes |
Crack (1135); Unintended Ejection (1234); Failure to Unfold or Unwrap (1669); Defective Component (2292); Failure to Advance (2524)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A health care professional reported that during an intraocular lens (iol) implant procedure, it was noted that the lens does not seem to be opening properly, was going in upside down, the haptic was breaking off and the lens had a crack in it.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information provided in h.6.(fda product problem code - 1234) the manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A health care professional reported that during an intraocular lens (iol) implant procedure, it was noted that the lens does not seem to be opening properly, was going in upside down, the haptic was breaking off.Additional information was received stating that the surgery was completed using another lens.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information provided in h.6.Correction: on initial mdr the product problem code of 1135 was an error.The lens was not returned for evaluation.Only the device was returned loose in the carton.The plunger lock and lens stop were removed.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger was advanced to mid-nozzle.No damage was observed.Complaint and product history records were reviewed and documentation indicated the product met release criteria.A qualified viscoelastic was indicated.The root cause could not be determined for the reported complaint of lens went in upside down.The reported broken haptic could not be verified because the iol was not returned for an evaluation.The ifu instructs: after the lens has been advanced to the ¿fill-to¿ line, the lens should be visually inspected to determine the position of the haptics.The plunger should be in contact with the trailing optic edge.No part of the lens should exit the nozzle prior to insertion through the incision.Once the lens is in position at the ¿fill-to¿ line, it should be implanted within 3 minutes.Proceeding with implantation of a misfolded haptic or a lens that appears to be ¿out of position¿ can result in a broken haptic or other negative outcome, since the haptic may be trapped and stretched, and/or pinched and sheared by the moving plunger.The ifu also instructs: if the leading haptic is straight or looped and extended in front of the lens, rotate device clockwise to be bevel left before advancing plunger to ensure the leading haptic is correctly placed in the capsular bag.As the leading haptic begins to exit the nozzle tip, place the leading haptic into the capsular bag in its correct orientation, and continue to slowly advance the plunger to deliver the lens.As the optic exits the nozzle, rotate the device counter clockwise back to center or slightly bevel right if needed to ensure the lens unfolds anterior side up within the capsular bag.Broken haptics may occur: if the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.In addition, haptic strength (modulus) decreases as the temperature increases and is more likely to break under stress.If a straight trailing haptic occurs and it was not properly detected to be out of position.If the plunger is not fully advanced, or if the plunger is allowed to retract, the trailing haptic may not release properly from the device.Any of the above listed causes alone, or in combination, may create the reported event.The manufacturer internal reference number is: (b)(4).
|
| |
|