| Catalog Number |
1410MSK |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiry date: 08/2025) h11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
|
| |
|
Event or Problem Description
|
|
It was reported that during an ultrasound guided breast biopsy procedure through normal density tissue, the needle of the device was allegedly bent.It was further reported that the needle was allegedly pulled with force from the patient's breast.A coaxial was used and procedure was completed with another device.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that during an ultrasound guided breast biopsy procedure through normal density tissue, the needle of the device was allegedly bent.It was further reported that the needle was allegedly pulled with force from the patient's breast.A coaxial was used and procedure was completed with another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the alleged needle bent and difficult to remove issues could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiry date: 08/2025), g3, h6 (method).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|