Catalog Number CXA260005
Medical Device Problem Codes Malposition of Device (2616); Migration (4003)
Health Effect - Clinical Codes Obstruction/Occlusion (2422); Reduced Blood Flow (4598)
Date of Event 11/09/2023
Type of Reportable Event Serious Injury
Event or Problem Description
The following information was reported to gore: on (b)(6) 2023, this patient underwent an endovascular treatment for an abdominal aortic aneurysm using gore® excluder® aaa endoprostheses, gore® excluder® conformable aaa endoprostheses and gore® dryseal flex introducer sheaths as accessories.The contralateral leg endoprosthesis was deployed in the patients right side/contralateral side.The leg could not be pushed up enough to the target site and unintentionally covered the bifurcation of the right internal iliac artery, resulting in reduced blood flow to the right internal iliac artery.Although pushing up the leg by using the edge of the 12 fr sheath improved blood flow, a dissection occurred by the sheath edge and blood flow into the false lumen of the right common iliac artery was observed.As a treatment, an additional bare stent was implanted in the right external iliac artery and blood flow to the false lumen disappeared.An angiography confirmed delayed blood flow into the right internal iliac artery.When an aortic extender was deployed in the proximal side, it moved proximally and covered halfway of the renal artery.The patient tolerated the procedure and was decided to be monitored.The physician reported the selection of 10 cm leg was fine, but device push-up during deployment was not successful.The physician considered that pushing up the leg by the edge of the 12 fr sheath was risky, but as the patient¿s left internal iliac artery was coil embolized, blood flow preservation to the right internal iliac artery was a priority.Reportedly the patient¿s vessels were in poor condition including calcification.
 
Additional Manufacturer Narrative
H.6: code c19: a review of manufacturing records verified that the lot involved in this complaint met all pre-release specifications.H.6: code b20: device remains implanted and therefore not available for direct analysis.H.6.: code d12: according to the gore® excluder® conformable aaa endoprosthesis instructions for use (ifu), potential adverse events that may occur and/or require intervention include, but are not limited to, renal artery occlusion and improper component placement.H.6.: code d1102: according to the gore® excluder® conformable aaa endoprosthesis instructions for use, implant procedure: do not continue advancing any portion of the delivery system if resistance is felt during advancement of the guidewire, sheath, or catheter.Stop and assess the cause of resistance.Vessel or catheter damage or premature deployment have occurred and may result in potential patient harms.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL SILICON VALLEY B/P
2890 de la cruz blvd.
santa clara CA 95050
Manufacturer Contact
kathryn irwin
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key18273859
Report Number3013164176-2023-01921
Device Sequence Number17528066
Product Code MIH
UDI-Device Identifier00733132650941
UDI-Public00733132650941
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P200030
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/16/2026
Device Catalogue NumberCXA260005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2023
Initial Report FDA Received Date12/06/2023
Date Device Manufactured05/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;