• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SP*2 POLY TIB COMP IMPACTOR; KNEE INSTRUMENT : IMPACTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SP*2 POLY TIB COMP IMPACTOR; KNEE INSTRUMENT : IMPACTORS Back to Search Results
Catalog Number 966385
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/01/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the impactor is broken.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).Investigation summary: according to the information received, it was reported that broken impactor.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.The physical investigation revealed that the poly impaction was broken in 2 pieces.The broken part was returned.The observed condition was identified as an end-of-life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since the broken condition.The overall complaint was confirmed as the observed condition of the sp*2 poly tib comp impactor would contribute to the complained device issue.Based on the investigation findings, potential cause can be attributed to end of life and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code (5/15/2005) provided is not a valid finished goods lot#.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
KNEE INSTRUMENT : IMPACTORS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18274149
Report Number1818910-2023-24665
Device Sequence Number7046727
Product Code HWA
UDI-Device Identifier10603295248279
UDI-Public10603295248279
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number966385
Device Lot NumberJ0505
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2023
Supplement Date Received by Manufacturer01/18/2024
02/20/2024
Initial Report FDA Received Date12/06/2023
Supplement Report FDA Received Date01/18/2024
02/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/15/2005
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-