Product complaint # = > (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Product complaint # (b)(4).Investigation summary: according to the information received, "instrument broke yesterday during knee surgery.The attachment end where the t-handle attaches is broken it doesn¿t allow the t-handle to be attached." the product was not returned to depuy synthes, however photos were provided for review.See attachment (b)(4) damaged im rod 400mm.The photo investigation revealed that the device 966120, sp2 im rod 400mm provided evidence was not sufficient to confirm the reported event and functionality issues can not be evaluated through a photo investigation.No further evidence can be observed from the photo to determine the root cause.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was unconfirmed as the observed condition of the 966120, sp2 im rod 400mm would not contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the device lot number is unknown, therefore a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.
|