| Catalog Number |
107140 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported a blood leak/coagulated blood was observed originating from the luer connection between the prismaflex hf1000 set and the thermax blood warmer disposable during continuous renal replacement therapy.The blood remained coagulated around the connection site and no blood was noted on the floor.The volume of blood loss was not reported.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the device was received for evaluation.Visual inspection of the set showed blood leakage between the luer warmer connection and the thermax bag.Functional testing was not performed on the actual sample as information regarding contamination was not obtained.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|