As reported in a research article, a 10mm amplatzer vascular plug 2 was placed off label to close a patent ductus arteriosus and was noted to protrude into the aorta so the device was removed.A more comprehensive assessment could not be performed as the event was non-contemporaneously reported through a literature review and no device was received for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
|
|
The article, ¿risk factors associated with device embolisation or malposition during transcatheter closure of patent ductus arteriosus¿, was reviewed.The article presented a case study of a 2 month old patient weighing 3.7 kg.It was reported that on an unknown date, an 8mm amplatzer vascular plug ii was chosen for implant to treat patent ductus arteriosus (pda) that was sized 6 mm.It was then reported on an unknown date, the device had embolized and a decision was made to retrieve the device via transcatheter snare and replace with an 10mm amplatzer vascular plug ii.During the implant procedure with the 10mm amplatzer vascular plug ii, it was noted the device led to aortic protrusion and a decision was made to remove the device.It is unknown if the second device was released from the delivery cable prior to removal from the patient.The article concluded device embolisation complicates 4% of transcatheter duct closures, with need for surgery in one-fifth of them.Larger ducts with high pulmonary artery pressures in younger and smaller infants are more often associated with device embolisation.Tubular ducts are more prone for embolisation compared to usual conical ducts.Softer vascular plugs are often associated with embolisations.Intentional device undersizing to avoid vascular obstruction in small patients is a frequent risk factor for embolisation.Precise echocardiographic measurements, correct occluder choice, proper technique and additional care in patients with high pulmonary artery pressures are mandatory to minimise embolisations.[the primary and secondary author was kothandam sivakumar, madras medical mission, 4a dr jj nagar, mogappair, chennai 600089, india, with corresponding email drkumarsiva@hotmail.Com].
|