EXACTECH, INC. SPECIFIC DEVICE NOT REPORTED; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
|
Back to Search Results |
|
| Catalog Number |
320-36-03 |
| Medical Device Problem Codes |
Unstable (1667); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Joint Dislocation (2374); Joint Laxity (4526)
|
| Date of Event |
11/09/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, approximately 3 months post op initial right tsa, this 64 y/o female patient was revised.Patient dislocated and became unstable.Surgeon removed the small proximal body, +5 humeral adapter tray, +2.5mm 36mm liner.To create greater stability the surgeon added a 25mm hrp middle segment, +0 proximal humeral body small, +0mm humeral adapter tray, 40mm +0 constrained humeral liner.Surgeon went from a 36mm djo glenosphere to a 40mm glenosphere.Patient was last known to be in stable condition following the event.Product not returning - hospital policy.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (h3) pending evaluation (d10) concomitant device(s): a487204 - 308-01-09 - 9x80mm distal stem modular cemented.(b)(6) - 308-05-17 - distal fixation ring ha 17.5.(b)(6) - 308-09-00 - small prox body +0.(b)(6) - 308-10-25 - 25mm middle segment modular.(b)(6) - 308-10-75 - 75mm middle segment modular.(b)(6) - 308-16-00 - taper locking screw 100.(b)(6) - 320-10-05 - equinoxe reverse tray adapter plate tray +5.(b)(6) - 320-20-00 - eq reverse torque defining screw kit.(b)(6) - 320-36-03 - 145-deg pe 36mm hum liner +2.5.
|
| |
|
Additional Manufacturer Narrative
|
|
The cause of the patient's shoulder dislocation and instability and subsequent revision reported cannot be conclusively determined; however, it may be due to soft tissue imbalance, patient anatomy, implant positioning, and/or implant selection.Dislocation and instability are known risks, as outlined in the ifu.D1: corrected.D4/d6: device, serial #, udi #, expiration, implant and explant dates unknown.G3: manufactured dates unknown.G4: pma 510k cannot be determined.H6: corrected health effect, medical device, component, and investigation clinical codes.
|
| |
|
Search Alerts/Recalls
|
|
|