MAUDE Adverse Event Report: MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 5.5PCF
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code Unintended Extubation (4564)
Date of Event 10/24/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the tracheostomy tube's cuff had leak during used, and  was replaced by the home care nurse.
 
Additional Manufacturer Narrative
Additional information: h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the cuff was burst and the inflation line was broken at the flange connection.The inflation line was partially broken at the connector flange joint and part of the inflation line tubing remained bonded into the flange.Functionally, an inflation/deflation test was performed using a syringe.Air was applied to the cuff, and it was observed that the cuff deflated immediately.It was reported that the tracheostomy tube's cuff had a leak during use.The reported issue was confirmed.The most likely causes were determined to be manufacturing related in regards to the broken inflation line and the device not being used as intended in relation to the burst cuff.Internal process improvements have been initiated to mitigate the issue of the broken inflation line at the flange connection.The instructions included with this device provide the following guidance: each tube¿s cuff, pilot balloon and valve should be tested by inflation prior to tube insertion.If dysfunction is detected in any part of the inflation system, the tube should not be used.Do not overinflate the cuff.Ordinarily, cuff pressure should not exceed 25 cm of h2o.Closely monitor cuff pressure under a physician¿s guidance.Overinflation may lead to permanent tracheal damage, rupture of the cuff with subsequent deflation, or cuff distortion that may cause airway blockage.To ease insertion and to guard against cuff perforation from sharp edges of cartilage, the cuff should be pulled back.Inflate the cuff and gently move the cuff away from the distal tip of the cannula towards the neck plate as the residual air is removed by deflation.Do not use sharp instruments such as forceps or hemostats that would damage the cuff when manipulating it.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX  32590
Manufacturer (Section G)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX   32590
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18279052
Report Number2936999-2023-01134
Device Sequence Number20500009
Product Code JOH
UDI-Device Identifier10884521103566
UDI-Public10884521103566
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K182861
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5.5PCF
Device Catalogue Number5.5PCF
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2023
Supplement Date Received by Manufacturer03/14/2024
Initial Report FDA Received Date12/06/2023
Supplement Report FDA Received Date04/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-