| Catalog Number |
P2000 |
| Medical Device Problem Code |
Power Problem (3010)
|
| Health Effect - Clinical Code |
Thrombosis/Thrombus (4440)
|
| Date of Event |
11/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that patient experienced minor complication of blood clots.There was no mention of any significant delay or tissue damage created by the transient development of blood clots in the irrigant.Once bladder irrigation flow was restored, the blood clots were reported to resolve.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: power shut off the failure identified in the investigation is inconsistent with the alleged failure.Probable root cause: device not plugged into a dedicated 20a outlet.Per ifu the p2000 requires a dedicated 20a outlet.The reported failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that patient experienced minor complication of blood clots.There was no mention of any significant delay or tissue damage created by the transient development of blood clots in the irrigant.Once bladder irrigation flow was restored, the blood clots were reported to resolve.
|
| |
|