Model Number N/A
Medical Device Problem Codes Component Incompatible (1108); Material Frayed (1262)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2023
Type of Reportable Event Malfunction
Event or Problem Description
Customer reported after the guidewire enters the puncture needle, it cannot enter the blood vessel, and the needle and guidewire are withdrawn, and the guidewire cannot be removed from the puncture needle.No other information provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and is pending evaluation.Results are expected soon.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
Customer reported after the guidewire enters the puncture needle, it cannot enter the blood vessel, and the needle and guidewire are withdrawn, and the guidewire cannot be removed from the puncture needle.No other information provided.
 
Event or Problem Description
Customer reported after the guidewire enters the puncture needle, it cannot enter the blood vessel, and the needle and guidewire are withdrawn, and the guidewire cannot be removed from the puncture needle.No other information provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a damaged guidewire was confirmed; however, the root cause was not identified.The product returned for evaluation was two photographs which depicted one stainless steel j-tip guidewire.The wire was inserted into its plastic hoop.The distal end of the wire was visible protruding from the hoop.The core wire was broken and the coil wire was elongated.The weld tip was present and remained attached to the coil wire.Damage was evident in the photographs; however, inspection of the photographs was insufficient to identify the cause of the damage.Consequently, this complaint is confirmed as ¿cause unknown¿ at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIAGARA SLIM-CATH DIALYSIS CATHETER KIT 12F X 20CM (STRAIGHT) (SHORT-TERM) DL
Common Device Name
CATHETER, HEMODIALYSIS, NON-IMPLANTED
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
suedabba mahboobi
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key18279801
Report Number3006260740-2023-05545
Device Sequence Number20544622
Product Code MPB
UDI-Device Identifier00801741045363
UDI-Public(01)00801741045363
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K010778
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number5553200
Device Lot NumberREGQ1654
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2023
Supplement Date Received by Manufacturer01/05/2024
02/16/2024
Initial Report FDA Received Date12/06/2023
Supplement Report FDA Received Date01/11/2024
02/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-