| Catalog Number |
06368590190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The cobas e411 disk serial number was (b)(6).The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable anti-tpo elecsys results from the cobas e411 disk analyzer.Patient 1 initial result was 152.5 iu/ml with a data flag.The repeat result on another cobas e411 was 50.68 iu/ml with a data flag and reported outside of the laboratory.The repeat result on the original analyzer was 48.68 iu/ml with a data flag.Patient 2 initial result was 151.2 iu/ml with a data flag.The repeat result on another cobas e411 was 11.54 iu/ml and reported outside of the laboratory.The repeat result on the original analyzer was 13.57 iu/ml.Patient 3 initial result was 341.7 iu/ml with a data flag.The repeat result on another cobas e411 was 19.91 iu/ml and reported outside of the laboratory.The repeat result on the original analyzer was 23.20 iu/ml with a data flag.Patient 4 initial result was 131.4 iu/ml with a data flag.The repeat result on another cobas e411 was 59.11 iu/ml with a data flag and reported outside of the laboratory.The repeat result on the original analyzer was 69.04 iu/ml with a data flag.
|
| |
|
Additional Manufacturer Narrative
|
|
The field service engineer replaced micro-particle stirrer (bead mixer) due to the higher generation of foam in the reagent.The investigation determined the service actions resolved the issue.
|
| |
|