LIMACORPORATE S.P.A. SMR SHOULDER; ECCENTRICAL GLENOSPHERE DIA. 36 MM Back to Search Results
Model Number 1376.09.031
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Hypersensitivity/Allergic reaction (1907); Implant Pain (4561)
Date of Event 10/11/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery performed on 11th october 2023 due to pain and nickel-metal allergy.The patient's shoulder prosthesis has been revised with the compassionate use device cmd 23-1224.The components involved in the revision surgery were the following: - smr reverse humeral body (product code 1352.15.010, lot 2214887, sterilization number 2200196) - smr reverse liner standard (product code 1360.50.810, lot 20at4ux, sterilization number 2100137) - smr eccent.Glenosphere ø 36mm (product code 1376.09.031, lot 2217722, sterilization number 2200224) - tt hybrid reverse baseplate #s (product code 1379.15.160, lot 2209111, sterilization number 2200136) the previous surgery was performed on december 5th 2022.Event happened in united states.
 
Additional Manufacturer Narrative
The dhrs of the involved product codes and lot numbers have been checked, without finding any anomaly.A final report will be submitted once the investigation is complete.
 
Event or Problem Description
Revision surgery performed on (b)(6) 2023 due to pain and nickel-metal allergy.The patient's shoulder prosthesis has been revised with the compassionate use device cmd 23-1224 (titanium glenosphere).The components removed in the revision surgery were the following: smr reverse humeral body (part code 1352.15.010, lot number 2214887, sterilization (b)(4).Smr reverse liner standard (part code 1360.50.810, lot number 20at4ux, sterilization (b)(4).Smr eccent.Glenosphere ø 36mm (part code 1376.09.031, lot number 2217722, sterilization (b)(4).There components were replaced by the following new devices: smr reverse humeral body (part code 1352.15.010, lot number 2319433, sterilization (b)(4).Smr reverse liner standard (part code 1360.50.810, lot number 23at0vs, sterilization (b)(4).Smr ecc.Glenosphere ø 36mm (part code 1376.15.031, lot number 2310423, sterilization (b)(4).This component was made available as compassionate use device.The previous surgery was performed on (b)(6) 2022.This was another revision surgery due to cuff failure.The patient is a female, date of birth (b)(6) 1960.Event happened in united states.
 
Additional Manufacturer Narrative
Investigation the dhrs of the involved product codes and lot numbers have been checked, without finding any pre-existing anomaly, that could have contributed to the event.This is the first and only complaint received on these lot numbers.We did not receive any radiography or explanted components.However, we received from the complaint source the following information: "the patient had a revision of the smr cocr glenosphere for a titanium one, due to being positive for a nickel allergy on 11th october 2023.The glenosphere was successfully replaced, as well as the humeral body and liner that was previously implanted during a surgery that occurred on 5th december 2022.There was no reason provided for replacing these components nor was there any issue reported by the surgeon.The humeral stem was noted to be well fixed and retained as was the glenoid baseplate.Additionally, the agent confirmed on 21st november 2023 that a new baseplate (code 1379.15.160, lot number 2201696) was not implanted.The surgeon simply used the safety screw from the baseplate box to secure the new glenosphere.Postoperatively, the patient expressed feeling much better than before, and no issues were reported." therefore, considering that: no pre-existing anomaly was discovered by checking the manufacturing charts of the components involved in this event the reason for the revision was mainly related to the nickel-metal allergy of the patient.We can conclude that the event is not product related.Pms data according to the available pms data, the revision rate of smr glenospheres belonging to the family product codes 1374.09.Xxx, 1376.09.Xxx, 1374.15.Xxx, 1376.15.Xxx due to allergy is lower than 0.01%.Based on the root cause analysis performed and according to the relevant pms data, no corrective actions are required for this specific case.The manufacturer will continue monitoring the market to promptly detect any further similar issues.Note: this is a final mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMR SHOULDER
Common Device Name
ECCENTRICAL GLENOSPHERE DIA. 36 MM
Manufacturer (Section D)
LIMACORPORATE S.P.A.
via nazionale 52
villanova di san daniele, udine 33038
IT  33038
Manufacturer Contact
via nazionale 52
villanova di san daniele, udine 33038
MDR Report Key18283678
Report Number3008021110-2023-00134
Device Sequence Number5909882
Product Code MBF
UDI-Device Identifier08033390087834
UDI-Public08033390087834
Combination Product (Y/N)N
PMA/510(K) Number
K163397
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model Number1376.09.031
Device Lot Number2217722
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/07/2023
Supplement Report FDA Received Date09/15/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-