• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTERLOCK; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTERLOCK; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 83779
Medical Device Problem Codes Stretched (1601); Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/05/2023
Type of Reportable Event Malfunction
Event or Problem Description
Reportable based on device analysis completed on 13nov2023.It was reported that there was obvious resistance, and the coil could no longer be pushed after about half of its release.The target lesion was located in the splenic artery.An 8mm x 20cm interlock was selected for use.During the procedure, it was noted that there was obvious resistance, and the coil could no longer be pushed after about 1/2 of its release.The coil was withdrawn, and it was noted that the coil was stretched.The procedure was completed with another of the same device.No complications were reported, and patient was stable post procedure.However, device analysis revealed that the main coil was detached.
 
Additional Manufacturer Narrative
E1 - initial reporter facility name: (b)(6).E1 - initial reporter address 1: (b)(6).E1 - initial reporter address 2: (b)(6).Device evaluated by mfr: only the main coil was returned for the analysis, and it was observed that the main coil was stretched, bent and detached at the middle section.The pouch was returned, and it was observed that the pouch information matches with the complaint information.No more damages were observed.The functional could not be performed due only the main coil was returned for the analysis.Dimensional inspection of the main coil revealed the components were within specification.No other issues were identified during the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLOCK
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18284044
Report Number2124215-2023-64583
Device Sequence Number17689599
Product Code KRD
UDI-Device Identifier08714729765011
UDI-Public08714729765011
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/29/2025
Device Model Number83779
Device Catalogue Number83779
Device Lot Number0029311613
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/13/2023
Initial Report FDA Received Date12/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MICROCATHETER: EPTECH
Patient Age52 YR
Patient SexFemale
Patient Weight55 KG
-
-