EXACTECH, INC. NV GXL LNR, NEUTRAL, 32MM ID, GROUP 3 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Catalog Number |
130-32-53 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Cyst(s) (1800); Failure of Implant (1924); Osteolysis (2377)
|
| Date of Event |
11/15/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Section h10: (h3) pending evaluation.(d10) concomitant device(s): 3627610 180-01-58 - crown cup,cluster-hole gr.58.
|
| |
|
Event or Problem Description
|
|
It was reported that this male patient was revised 8 years postop initial tha.The x-ray control showed a clear decentering of the prosthetic head, unusually large osteolysis in the acetabulum as a sign of inlay wear.This could happen to one.Inlay change on november 15, 2023 to a vitd-hardened, specially approved inlay.As part of the replacement operation, in addition to the inlay replacement, the solid cup integrity was determined as well as curettage and sealing the cysts in the socket using allogeneic spongiosa and changing the prosthetic head.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (h3) the prosthesis wear reported may have been due to a combination of the risk factors specified in an hhe such as use error, implant positioning, implant size selection, patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential), and/or surgical approach.However, this cannot be confirmed since the devices were not returned for evaluation and radiographs were not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: h6 - health effect clinical code, component code, investigation conclusions, investigation conclusions.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|