• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APOLLO ENDOSURGERY, INC -TACK¿ ENDOSCOPIC HELIX TACKING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APOLLO ENDOSURGERY, INC -TACK¿ ENDOSCOPIC HELIX TACKING SYSTEM Back to Search Results
Model Number XTACK-160-H
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 01/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Combined medwatch submitted to the fda on 07/dec/2023.Clinical study (b)(6).A review of the device labeling notes the following: the current x-tack¿ endoscopic helix tacking system instructions for use (ifu) addressed the known and potential events of "bleeding, hospitalization, medication, and blood transfusion " as follows: warnings - the device should not be used to treat acutely bleeding ulcers, ulcers with stigmata of recent bleeding or any ulcers with a visible vessel.- do not use a device where the integrity of the sterile packaging has been compromised or if the device appears damaged.- only physicians possessing sufficient skill and experience in similar or the same techniques should perform endoscopic procedures.- contact of electrosurgical components with other components may result in injury to the patient and/or operator as well as damage to the device and/or endoscope.- verify compatibility of endoscope size, endoscopic instruments and accessories and ensure performance is not compromised.- ensure endoscope is clean, dry, and free of lubricants prior to device installation.- ensure all endoscopic scopes, including scope channels, are in good working condition prior to use.- suction operation through endoscope may be significantly reduced when the scope channel liner is in proper position.- do not push through or pull back on a retroflexed scope with installed helix tack.- applying excessive force to the distal end of the x-tack¿ device could compress or damage the helix tack when installed.- do not retract device into scope whilst a helix tack is installed.- reuse or reprocessing of the x-tack¿ system could result in device malfunction, precautions - the system may only be used if purchased from apollo endosurgery, inc.Or one of its authorized agents.Adverse events possible complications that may result from using the x-tack¿ system include, but may not be limited to: - pharyngitis / sore throat - nausea and / or vomiting - abdominal pain and / or bloating - hemorrhage - hematoma - conversion to laparoscopic or open procedure - stricture - infection / sepsis - pharyngeal, gastric, colonic and/or esophageal perforation - esophageal, gastric, colonic and/or pharyngeal laceration - intra-abdominal (hollow or solid) visceral injury - aspiration - wound dehiscence - acute inflammatory tissue reaction - death additional information: the device has not been returned for analysis.The investigator determined a device history record (dhr) review is not possible for this complaint, as the reporter noted the device lot number was unavailable.Device evaluation summary: assessment of the device involved in this complaint was not possible, and it has not been possible to determine the root cause for this event.
 
Event or Problem Description
This report was report from a clinical study.On the follow up visit the patient presented with bleeding.Patient was admitted to the hospital where she was had and endoscopy performed and it was found patient was bleeding.The patient was giving medication and blood transfusion.She was discharged 4 days after being admitted with no concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
-TACK¿ ENDOSCOPIC HELIX TACKING SYSTEM
Common Device Name
ENDOSCOPIC HELIX TACKING SYSTEM
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC
1120 s. captail of texas hwy
bldg 1, ste 300
austin 78746
Manufacturer (Section G)
APOLLO ENDOSURGERY COSTA RICA, SRL
coyol free zone
building b 13.3
alajuela, cs CRI
CS   CRI
Manufacturer Contact
adriana russell
1120 s. captail of texas hwy
bldg 1, ste 300
austin 78746
5122795114
MDR Report Key18284979
Report Number3006722112-2023-00230
Device Sequence Number6696038
Product Code OCW
UDI-Device Identifier10811955020763
UDI-Public(01)10811955020763
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K201808
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberXTACK-160-H
Device Catalogue NumberXTACK-160-H
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/17/2023
Initial Report FDA Received Date12/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other;
Patient Age79 YR
Patient SexFemale
Patient Weight42 KG
-
-