• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERLOC SAFETY INFUSION SET 19G X 1 IN WITH Y-SITE; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERLOC SAFETY INFUSION SET 19G X 1 IN WITH Y-SITE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Chemical Exposure (2570)
Date of Event 11/13/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by customer that no patient harm, but patient was exposed to chemo leak and was in public when the leak occurred.The connector came off of the tubing exposing the lumen.Patient called, advised to turn off and come in to be evaluated the same day.A new cassette was given the next day to patient to receive medication.It was stated there was a delay in care.No other information was provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and is pending evaluation.Results are expected soon.
 
Additional Manufacturer Narrative
Section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a leak is unconfirmed and could not be reproduced.One 19g x 1" powerloc infusion set was returned for evaluation.The sample did not show signs of damage to the luers or tubing and was within specification of all parts.The sample was pressurized and a leak was not observed.Due to the complaint being unable to be reproduced, this complaint is unconfirmed.This complaint will be recorded for future trending and monitoring purposes.
 
Event or Problem Description
It was reported by customer that no patient harm, but patient was exposed to chemo leak and was in public when the leak occurred.The connector came off of the tubing exposing the lumen.Patient called, advised to turn off and come in to be evaluated the same day.A new cassette was given the next day to patient to receive medication.It was stated there was a delay in care.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERLOC SAFETY INFUSION SET 19G X 1 IN WITH Y-SITE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city UT 84116
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key18285617
Report Number3006260740-2023-05575
Device Sequence Number20544705
Product Code FPA
UDI-Device Identifier00801741047893
UDI-Public(01)00801741047893
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K060812
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number0671910
Device Lot NumberASHTFC025
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2023
Supplement Date Received by Manufacturer01/23/2024
Initial Report FDA Received Date12/07/2023
Supplement Report FDA Received Date02/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-