MAUDE Adverse Event Report: LAKE REGION MEDICAL SAFARI2¿; WIRE, GUIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LAKE REGION MEDICAL SAFARI2¿; WIRE, GUIDE Back to Search Results
Model Number H74939407XS0
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/27/2023
Type of Reportable Event Malfunction
Event or Problem Description
Event description: it was identified during the tavi procedure that the guide showed a quality deviation the site clarified that they identified the defect when opening the material.The issue occured prior to use.The procedure was completed using a different device (same model).No patient injury.
 
Additional Manufacturer Narrative
This medwatch report is being filed based on the image received on november 30, 2023.The image provided by the distributor presented the guidewire protruding from the dispenser hoop with a portion of the distal tip of the guidewire inside tubing.The guidewire presented stretched coil wraps and a fracture of the core wire material at an unknown distance from the distal tip.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.During manufacturing and packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.As described in the device instructions for use (dfu): 1.Ensure a catheter is properly placed in the ventricle or other intended treatment area.2.Carefully remove the safari2tm guidewire from the hoop by grasping the proximal end of the j-straightener separating the straightener from the hoop.3.After inspection of the safari2tm guidewire, advance the j-straightener over the distal section of the safari2tm guidewire to straighten the curve.4.Insert the safari2tm guidewire into the hub of the catheter with the aid of the j-straightener.5.Carefully advance the distal tip of the safari2tm guidewire into the lumen of the catheter.6.Remove the safari2tm guidewire j-straightener by withdrawing it over the length of the safari2tm guidewire.7.Advance the safari2tm guidewire through the catheter under fluoroscopic guidance.At this time, it is not possible to assign a definitive root cause for the event as reported.If there is any further relevant information provided, a follow up medwatch report will be submitted.
 
Additional Manufacturer Narrative
Device evaluation: as received, the specimen consisted of one-1 each safari2 275cm xsml crv 1p; returned coiled, loose without the dispenser assembly and double-bagged within "zip-lock" style poly biohazard pouches.Note, the distal tip was lodged within the j-straightener.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.During manufacturing and packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.The specimen presented a ductile, tensile overload of the core wire an unknown distance from the distal tip weld mass with concurrent stretched coil damage of 17.2 cm, the distal tip is lodged within the lumen of the j-straightener attachment 3.6 cm from the distal tip of the j-straightener along with traces of dried blood.The specimen also presented coil misalignment from 3 cm to 5.2 cm from the proximal aspect of the formed curve.Our investigation was unable to confirm that the product did not meet specification prior to shipment.The investigation concluded that the product met specification at the time of shipment.As described in the device instructions for use (dfu): 1.Ensure a catheter is properly placed in the ventricle or other intended treatment area.2.Carefully remove the safari2tm guidewire from the hoop by grasping the proximal end of the j-straightener separating the straightener from the hoop.3.After inspection of the safari2tm guidewire, advance the j-straightener over the distal section of the safari2tm guidewire to straighten the curve.4.Insert the safari2tm guidewire into the hub of the catheter with the aid of the j-straightener.5.Carefully advance the distal tip of the safari2tm guidewire into the lumen of the catheter.6.Remove the safari2tm guidewire j-straightener by withdrawing it over the length of the safari2tm guidewire.7.Advance the safari2tm guidewire through the catheter under fluoroscopic guidance.At this time, it is not possible to assign a definitive root cause for the event as reported.Based on the information provided and the evidence presented, it appeared that handling factors have contributed to the event as reported.If there is any further relevant informaiton provided, a follow up medwatch report will be submitted.
 
Additional Manufacturer Narrative
Correction to a previously submitted mdr to correct the primary unique device identifier (udi) number in section d4.
 
Additional Manufacturer Narrative
Correction to a previously submitted mdr to correct the brand name in section d1 and the catalog number in section d4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFARI2¿
Common Device Name
WIRE, GUIDE
Manufacturer (Section D)
LAKE REGION MEDICAL
340 lake hazeltine drive
chaska MN 55318
Manufacturer (Section G)
LAKE REGION MEDICAL
340 lake hazeltine drive
chaska MN 55318
Manufacturer Contact
sharon seifert
340 lake hazeltine drive
chaska, MN 55318
9526418518
MDR Report Key18286644
Report Number2126666-2023-00081
Device Sequence Number15144219
Product Code DQX
UDI-Device Identifier08714729887638
UDI-Public(01)08714729887638(17)250925(10)7376189
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K151244
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/25/2025
Device Model NumberH74939407XS0
Device Catalogue NumberH74939407XS0
Device Lot Number7376189
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/30/2023
Supplement Date Received by Manufacturer11/30/2023
Not provided
Not provided
Initial Report FDA Received Date12/07/2023
Supplement Report FDA Received Date02/23/2024
06/13/2024
06/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-