| Catalog Number |
5C4480C |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Puncture/Hole (1504); Protective Measures Problem (3015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that there were holes in an extension set which resulted in a leak that led to a system error 2240 (air in line/set) alarm.The patient was connected at the time of the alarm.This occurred during drain two of five of peritoneal dialysis (pd) therapy.Renal therapy services assisted the patient with ending the therapy session and reviewed proper procedures per the user manual.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|