| Model Number |
PM3542 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Capturing Problem (2891)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information requested, but not received.
|
| |
|
Event or Problem Description
|
|
It was reported.A patient's pacemaker presented with high capture thresholds and premature battery depletion.The device was explanted and replaced in a procedure on (b)(6) 2023.Post-procedure the patient was stable.
|
| |
|
Event or Problem Description
|
|
Related manufacturer reference number: 2017865-2024-00102.It was reported a patient's pacemaker presented with high capture thresholds and premature battery depletion.The left ventricular (lv) lead also had high capture thresholds.The pacemaker and lv lead were explanted and replaced in a procedure on (b)(6) 2023.Post-procedure the patient was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Field event of capture problem and premature depletion were not confirmed.As received device was in normal range of operation and no anomalies were noted.Output, pacing and capture functions of the device were tested and found to be normal.Further analysis performed, including electrical, thermal and mechanical testing of the device did not find any anomaly.Assessment of total longevity was performed, and device was found to have appropriate remaining longevity.
|
| |
|