VENTEC LIFE SYSTEMS, INC VOCSN PATIENT BREATHING PACKAGE; VENTILATOR, CONTINUOUS, FACILITY USE
|
Back to Search Results |
|
| Model Number |
PEDIATRIC, ACTIVE, OXYGEN, BLUE |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to ventec that the clear film wall of a breathing circuit had detached from the blue polypropylene spiral portion of the patient circuit (pediatric, active, oxygen, blue) assembly.The reported issue was discovered during incoming inspection.There were no reports of patient involvement associated with the reported event.Section d4 model number, catalog number and lot code reflects the information known about the patient circuit.The udi was unavailable at the time of this submission, therefore section d4 udi is "unknown", and section h4 date of manufacturer will be left blank.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: ventec continues to investigate the reported issue.A follow-up report will be submitted when the investigation is complete as defined by 21 cfr 803.56.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information for supplemental medwatch report 001 ¿ section d4, model # and catalog # of the initial medwatch report stated: prt-00802-000.Section d4, model # and catalog # of the initial medwatch report should have stated: prt-00802-001.New information for supplemental medwatch report 001 ¿ h6: udi information for the patient circuit was confirmed.Section d4, unique identifier (udi) #, and section h4, date of manufacturer, have been updated accordingly.The patient circuit was returned to ventec.Ventec evaluated the patient circuit and observed that the inner clear tubing was delaminating (detaching) from the blue spiral reinforcement.The degree of delamination appeared to vary throughout the length of the circuit based on visual observation and tactile handling of the bond.Ventec contacted the contract manufacturer of the product (supplier) regarding the reported issue.The supplier later advised ventec that it had performed an internal investigation and determined that the product was manufactured using inadequate extrusion temperature.Please reference ventec corrections and removals report number 3013095415-12/22/2023-001-r.Ventec was unable to enter and transmit the corrections and removals number in section h9 due to system character limits in that field.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi related data quality updates only.Corrected information for supplemental medwatch report 002 ¿.Section d1, brand name, of the initial medwatch report stated: vocsn.Section d1, brand name, of the initial medwatch report should have stated: vocsn patient breathing package.Section d4, model #, of supplemental medwatch report 001 stated: prt-00802-001.Section d4, model #, of supplemental medwatch report 001 should have stated: pediatric, active, oxygen, blue.Section d4, udi #, of supplemental medwatch report 001 stated: (b)(4).Section d4, udi #, of supplemental medwatch report 001 should have stated: (b)(4).Section h5, labeled for single use, of the initial medwatch report stated: no.Section h5, labeled for single use, of the initial medwatch report should have stated: yes.
|
| |
|