Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: d9: complainant part is not expected to be returned for manufacturer review/investigation.E1: initial reporter facility name: (b)(6).H3, h4, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history review (dhr): part: 04.613.516s, lot: 6l97569, manufacturing site: werk selzach, supplier: (b)(4), release to warehouse date: 26 may 2020, expiration date: 01 may 2030.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Non-sterile part: part number: 04.613.516, lot number: 51p0121, manufacturing site: mezzovico, release to warehouse date: 11 may 2020.A manufacturing record evaluation was performed for the not sterile finished lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Device report from synthes reports an event in japan as follows: it was reported that on (b)(6) 2023, the patient underwent an anterior cervical fusion for ossification of the posterior longitudinal ligament.The surgery was completed successfully without any surgical delay.After the surgery, there was no problem on inspection in september 2022.The screws in question are currently loose.Also, the plate in question is floating a few millimeters forward from the vertebrae.The patient is scheduled for removal and re-fixation on (b)(6) 2023.The plate will be shortened by 2mm, and the screw will be increased by 2mm from 16mm to 18mm.The patient status was reported to be stable.This report involves one 4.0mm ti cerv self-retain scr slf-drlg/variable angle 16mm.This is report 2 of 5 for (b)(4).
|