| Model Number |
08058687190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The serial number of the customer's cobas pro c 503 analytical unit is (b)(6).The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter received a questionable trigl triglycerides result from one patient sample tested on the cobas pro c 503 analytical unit.The initial result was 271 mg/dl.The repeat result was 129 mg/dl.
|
| |
|
Additional Manufacturer Narrative
|
|
The field applications specialist (fas) reviewed the event and determined the event was due to a preanalytic issue at the customer site.The fas inspected 30 random patient sample tubes and noted that 12 patient sample tubes had fibrin and gel when the tubes were inverted.The investigation did not identify a product problem.
|
| |
|
Additional Manufacturer Narrative
|
|
**udi related data quality updates only** d3 is the initial distributor in the us.
|
| |
|