|
The reported device was received for evaluation.A visual inspection revealed that the returned device was not in its original packaging.The distal end of the screw was fractured away from the proximal body of the screw and broken into several pieces.There was biological debris on the returned fragments.An assessment of material characteristics found that based on the condition of the product material found during visual inspection, additional material testing is not required.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.The root cause has been associated with unintended use of the device.Factors that could have contributed to the failure include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
|