BARD PERIPHERAL VASCULAR, INC. CONQUEST; PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number CQ5074
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/19/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an angioplasty procedure in axillary vein, the pta balloon was allegedly ruptured after twelve second of expansion.The procedure was completed by using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records will be performed.The return of the sample is pending.However, photo was provided for review.The investigation of the reported event is currently underway.H10: d4 (expiry date: 08/2025).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
H10: additional information was received, the file was reassessed for reportability and determined to be no longer reportable.Since an initial mdr was submitted, therefore, the file will remain assessed as a malfunction.H10: d4 (expiry date: 08/2025), g3.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Device pending return.
 
Event or Problem Description
It was reported that during an angioplasty procedure in axillary vein, the pta balloon was allegedly ruptured after twelve second of expansion.The procedure was completed by using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: additional information provided by the complainant indicated that there was no vessel damage, no detachment of the device, and no difficulty retracting the device through the introducer sheath; therefore, this event can be deemed a non-complicated balloon rupture and is no longer reportable.However, our firm received an fda email correspondence on 8-july-2024 from mdr data systems team, (b)(4) indicating that the information (specifically the unique device identifier (udi) information in section d) of the previously submitted fda form 3500a, was incorrect and a request was made to submit supplemental report(s) with the entire udi number including the di and pi fields, all numbers, letters, parentheses, and symbols in the ¿unique device identifier (udi) #¿ fields.This supplemental report is in response to that request.H11: d4: medical device expiration date- 08/31/2025 d4: udi- (b)(4).Device manufacture date- 11/30/2022.
 
Event or Problem Description
It was reported that during an angioplasty procedure in axillary vein, the pta balloon allegedly ruptured after twelve seconds of expansion.The procedure was completed by using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key18298421
Report Number2020394-2023-01210
Device Sequence Number20575622
Product Code DQY
UDI-Device Identifier00801741062957
UDI-Public(01)00801741062957(11)221130(17)250831(10)REGW2739
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K083657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Catalogue NumberCQ5074
Device Lot NumberREGW2739
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/19/2023
Supplement Date Received by Manufacturer12/09/2023
07/08/2024
Initial Report FDA Received Date12/10/2023
Supplement Report FDA Received Date01/05/2024
07/19/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-