The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb as defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module with drive pcb.In addition, our technician confirmed that the us connector cable (long) buckled, the objective prism cracked, the ultrasound connector body corroded, the remote control buttons leak, the remote control buttons cut, the electrical pin connector dirty, the lg air/water socket cylinder dirty, the warning/caution label worn out, and the ccd module with drive pcb distorted image; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|