MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS SET: 2-LUMEN 12 FR X 16 CM; CATHETER PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS SET: 2-LUMEN 12 FR X 16 CM; CATHETER PERCUTANEOUS Back to Search Results
Catalog Number CS-12122-F
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that"(b)(6) 2023, the ars was found leak during used on the patient.The patient was reported as fine.
 
Additional Manufacturer Narrative
Qn#(b)(4).
 
Event or Problem Description
It was reported that"(b)(6) 2023, the ars was found leak during used on the patient.The patient was reported as fine.
 
Additional Manufacturer Narrative
(b)(4).The customer returned one guide wire assembly, one dilator, and one catheter over needle for analysis.Signs of use in the form of biological material were observed on the guide wire.It was noted that the arrow raulerson syringe (ars) nor the 18ga introducer needle involved with this complaint were returned for analysis.Visual inspection of the ars and introducer needle could not be performed as they were not returned for analysis.It was noted that the guide wire body was kinked.The customer was contacted to address this discrepancy, and they confirmed that they do not want to report on the kinked guide wire.Therefore, the circumstances why the guide wire was kinked cannot be verified.The module requirement document for raulerson syringes (amrq-000113 rev.03) was reviewed to determine requirements for air/water leakage.The document notes a deviation from iso 7886-1: "the freedom of air and liquid leakage past the piston requirement is design restrictive and is intended for an injection-intended syringe, not the ars.The opening in the center of the piston that allows passage of the inner cannula prohibits the ars from meeting the pressure and vacuum requirements as dictated by the standard.However, because the intended use of the ars is to allow aspiration of blood to ensure venous placement of the introducer needle and to aid in the insertion of the spring wire guide, the leakage requirements of a standard syringe are not applicable to the ars." despite this, additional testing cannot be performed on the ars as it was not returned for analysis.Device history record review was performed, and no relevant findings were identified.The ifu provided with the kit informs the user, "insert introducer needle or catheter/needle with attached syringe or arrow raulerson syringe (where provided) into vein and aspirate".The issue of ars leaking could not be confirmed as part of this complaint investigation.Visual and functional analyses could not be performed as neither the ars nor the introducer needle were returned for analysis.A device history record review was performed, and no relevant findings were identified.Based on the customer report and the sample received, the root cause cannot be determined at this time.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
It was reported that"on (b)(6) 2023, the ars was found leak during used on the patient.The patient was reported as fine.
 
Additional Manufacturer Narrative
(b)(4).In section d provided the full udi # udi related data quality updates only.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW HEMODIALYSIS SET: 2-LUMEN 12 FR X 16 CM
Common Device Name
CATHETER PERCUTANEOUS
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
marling heaton
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18299837
Report Number3006425876-2023-01217
Device Sequence Number15199512
Product Code DQY
UDI-Device Identifier00801902100108
UDI-Public(01)00801902100108(11)220816(17)270731(10)71F22H1708(21)
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K862056
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberCS-12122-F
Device Lot Number71F22H1708
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2023
Supplement Date Received by Manufacturer01/03/2024
Not provided
Initial Report FDA Received Date12/11/2023
Supplement Report FDA Received Date01/04/2024
07/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED; NOT REPORTED; NOT REPORTED
Patient SexUnknown
-
-